Magnetoencephalography with a chip-scale atomic magnetometer

Biomed Opt Express. 2012 May 1;3(5):981-90. doi: 10.1364/BOE.3.000981. Epub 2012 Apr 17.

Abstract

We report on the measurement of somatosensory-evoked and spontaneous magnetoencephalography (MEG) signals with a chip-scale atomic magnetometer (CSAM) based on optical spectroscopy of alkali atoms. The uncooled, fiber-coupled CSAM has a sensitive volume of 0.77 mm(3) inside a sensor head of volume 1 cm(3) and enabled convenient handling, similar to an electroencephalography (EEG) electrode. When positioned over O1 of a healthy human subject, α-oscillations were observed in the component of the magnetic field perpendicular to the scalp surface. Furthermore, by stimulation at the right wrist of the subject, somatosensory-evoked fields were measured with the sensors placed over C3. Higher noise levels of the CSAM were partly compensated by higher signal amplitudes due to the shorter distance between CSAM and scalp.

Keywords: (120.0120) Instrumentation, measurement, and metrology; (170.0170) Medical optics and biotechnology; (230.0230) Optical devices.