Abstract
Magnetoencephalography has long held the promise of providing a noninvasive tool for localizing epileptic seizures in humans because of its high spatial resolution compared with the scalp EEG. Yet, this promise has been elusive, not because of a lack of sensitivity or spatial resolution but because the large size and immobility of present cryogenic (superconducting) technology prevent long-term telemetry required to capture these very infrequent epileptiform events. To circumvent this limitation, we used Micro-Electro-Mechanical Systems technology to construct a noncryogenic (room temperature) microfabricated atomic magnetometer (“magnetrode”) based on laser spectroscopy of rubidium vapor and similar in size and flexibility to scalp EEG electrodes. We tested the magnetrode by measuring the magnetic signature of epileptiform discharges in a rat model of epilepsy. We were able to measure neuronal currents of single epileptic discharges and more subtle spontaneous brain activity with a high signal-to-noise ratio approaching that of present superconducting sensors. These measurements are a promising step toward the goal of high-resolution noninvasive telemetry of epileptic events in humans with seizure disorders.
Introduction
Neuronal currents associated with brain activity produce minute extracranial magnetic fields detectable in the magnetoencephalogram (MEG), accompanied by electrical potentials recorded in the scalp EEG. Over the past 30 years, there has been intense interest in MEG because, although electrical potentials are attenuated and distorted by the heterogeneous resistive properties of the cranium that compromise the spatial resolution of EEG, the skull is transparent to magnetic fields (Cohen and Cuffin, 1983). MEG has long presented the promise of providing noninvasive images of brain activity, with a spatial and temporal resolution rivaling that obtained only through penetrating the skull by use of electrodes placed within the brain.
Nowhere is the potential of MEG more clinically relevant than in the surgical treatment of human epilepsy (Barth et al., 1982, 1984), where the origin of seizures cannot always be adequately determined from the scalp EEG, making invasive intracranial electrical recording (electrocorticography [ECoG]) necessary (Engel, 2013). Yet, the use of MEG to replace invasive recording of epilepsy has remained elusive, primarily because present cryogenic technology is not suited to long-term telemetry required to capture single rare events (seizures). What is needed to fully realize the clinical potential of MEG are sensors that may be operated at room temperature, can be directly and flexibly attached to the scalp (e.g., EEG electrodes) for long-term telemetry, are of sufficiently small size to permit high density placement, and most important, are capable of detecting single epileptic discharges with a sensitivity similar to invasive recording electrodes. Here, we report recording single epileptiform spikes and smaller-amplitude spontaneous brain activity, in an animal model where MEG could be directly compared with ECoG, using a microfabricated atomic magnetometer (μAM) that meets these requirements.
Materials and Methods
μAM principle of operation.
Atomic magnetometers, also known as optically pumped magnetometers, use the influence of magnetic fields on alkali atoms in a vapor (see, e.g., Budker and Romalis, 2007). The vapor of atoms is spin-polarized by transferring the spin of resonant laser light to the atoms (Fig. 1A). Once the atoms (in our case rubidium) are polarized, a precession of the electron spin can be imparted by external magnetic fields (Fig. 1B). Reorientation of the spins changes the light transmitted through the atomic vapor. At zero magnetic field, the light transmission is maximum because most of the atoms are spin-polarized parallel to the transmitted light. In a nonzero transverse magnetic field, the atoms precess, and the static polarization, resulting from a balance between optical pumping and precession, will not be optimally aligned with the pump light, causing a decrease in the transmission intensity (Fig. 1C). To determine the direction of the magnetic field, a small oscillating magnetic field is applied perpendicular to the light beam, and the transmitted light is demodulated with a phase-sensitive detector.
Theory and operation of the μAM. A, Rubidium (87Rb) atoms are spin-polarized by transferring the spin of resonant laser light. B, Precession, or wobble, of electron spin is imparted by external magnetic fields, changing the light transmitted through the atomic vapor. C, Plot of light transmission with magnetic field strength (black; right) and the demodulated signal (gray; left). D, μAM (right) with photodiode removed to expose 87Rb vapor cell inside of vacuum package. The μAM is shown compared with a standard scalp EEG electrode (left) and grain of rice (top).
μAM construction.
We built a μAM (Schwindt et al., 2007) of comparable size to a standard scalp EEG electrode (Fig. 1D), using Micro-Electro-Mechanical Systems technology. The use of fiber-optics and evacuated packaging (Mhaskar et al., 2012; Sander et al., 2012) enabled flexible positioning, low-power operation, and the ability to position the heated sensor within a few millimeters of living tissue. The sensor head contained a 87Rb vapor cell within a vacuum package (Mescher et al., 2005) (Fig. 1D), keeping the outside of the sensor at room temperature while the cell was heated to 150°C. The thin vacuum gap allowed a very high-signal-to-noise ratio because the atomic vapor could be positioned within millimeters of the scalp. The vacuum package was glued to a microfabricated optical bench (Fig. 1D, underside of μAM) holding a ferrule with the optical fibers, along with a polarizer, beam splitter, mirror, and quarter-wave plate. The pump light emitted by one of the fibers was circularly polarized, transmitted through the rubidium cell, and collected on a photodiode (removed to expose vapor cell in Fig. 1D).
MEG measurement setup.
A distributed-feedback laser was used to pump and probe the atoms in four sensors simultaneously. The light was coupled into a polarization-maintaining fiber and split into four arms, one for each sensor, by means of a fiber splitter with manually adjustable attenuation. The heating light was provided by a second distributed-feedback laser and split into four channels with a fiber splitter. The heat power for each sensor was adjusted individually with variable optical attenuators. The signal from the photodiode was amplified and processed with a digital signal processor-based lock-in amplifier. A low-pass filter of eighth order at 400 Hz and notch filters at 60 Hz and 180 Hz were applied digitally.
The sensor heads were mounted in a 3D-printed plastic holder such that all fibers were pointing in the vertical direction. The holder was mounted to a Delrin platform. The platform was slid into a Plexiglas cylinder used as a manifold to hold a set of magnetic field coils to control the magnetic-field environment. The coil set could create three orthogonal magnetic fields homogeneous to 1:100 over a volume of 5 cm. In addition, a field gradient along the cylinder axis could be applied to the field along this axis. The coil manifold was housed in a three-layer magnetic shield made of μ-metal. The rat was positioned on a slider bed with a nonmagnetic 3D-printed nose clamp and two ear bars. The bed could be shifted in two horizontal directions under the sensor array, and the sensors could be lowered to touch the head of the animal. Two sensors were mounted in the vertical direction with a separation of 1 cm, and a third sensor was 10 cm away to monitor the background magnetic fields. All sensors measured the field component along the cylinder axis.
Animals and surgery.
All procedures were conducted within the guidelines established by the University of Colorado Institutional Animal Care and Use Committee. Adult male Sprague Dawley rats (n = 3; 510–515 g) were anesthetized to surgical levels using subcutaneous injections of ketamine (71 mg/kg of body weight), xylazine (14 mg/kg), and acepromazine (2.4 mg/kg). Additional doses were administered every 60 min to maintain an anesthetic depth where the corneal and flexor withdrawal reflexes could barely be elicited. Animals were placed on a regulated heating pad to maintain normal body temperature (37°C). Skin, fascia, and parts of the masseter muscle were removed to expose the dorsal and lateral aspects of the skull. Bilateral cranial windows (2 mm × 2 mm) were surgically opened on the lateral side of the skull at the junction of the squamosal and parietal bones by use of a small Dremel tool with a 1 mm burr tip under a surgical microscope. The dura was reflected and the exposed cortex regularly doused with Ringer's solution containing the following (in mm): 135 NaCl, 3 KCl, 2 MgCl2, and 2 CaCl2, pH 7.4, at 37°C. At the conclusion of the experiment, the animals were killed by anesthetic overdose without ever regaining consciousness.
Electrical recording.
Electrophysiology was performed using Ag wire ball-electrodes (ball diameter: ∼1 mm; bare wire diameter: 250 μm; Teflon-coated diameter: 330 μm). The electrodes were positioned epipially through the cranial window in the squamosal bone. Recordings were referenced to a similar electrode secured over the occipital bone and were amplified (×500; DAM80, World Precision Instruments), analog filtered (bandpass: 1–1000 Hz). The demodulated signals from the three μAMs and amplified field potentials from the ECoG electrodes were recorded with a 24-bit ADC system at a 10 kHz sampling frequency. Both MEG and ECoG signals were subsequently digitally bandpass filtered (1–100 Hz) using a zero-phase, second-order, Butterworth design. To evoke epileptiform spikes, the GABAA antagonist bicuculline methiodide (Tocris Bioscience; 100 μl, 4 mm, in 0.9% saline) was applied to the cortical surface subsequent to ketamine spike data collection. Epileptiform spiking from bicuculline methiodide application stabilized after 10 min, at which point data were collected in 5 min blocks.
Results
The intrinsic sensitivity of the μAM was measured to be <25 fT/Hz1/2 and averaged between 10 and 50 Hz. The noise from the magnetic shield was estimated to be ∼12 fT/Hz1/2, and the total sensitivity inside the shield can was at 30 fT/Hz1/2 during the experiment. The intrinsic bandwidth of the sensor was ∼250 Hz, rolling off similar to a first-order low-pass filter to a frequency set by the steep digital signal processor low-pass filter.
Figure 2 shows simultaneous recording from the μAM and from ECoG electrodes placed directly on the cortical surface of a rat. With the bottom of the μAM vacuum package pressed against the dorsal parietal skull, the center of the 87Rb cell was located ∼3.7 mm dorsal to the left parietotemporal ridge in Experiment 1 (Fig. 2A), or left and right parietotemporal ridges in Experiment 2 (Fig. 2D). In Experiment 1, an ECoG electrode recorded data from the pial surface (dura reflected) of the left hemisphere through a cranial window in the temporal bone 10 mm ventral to the 87Rb cell (Fig. 2A). In Experiment 2, bilateral ECoG monitoring was performed (Fig. 2D). The smallest signals recorded were spontaneous transient nonepileptiform discharges associated with ketamine anesthesia (Kayama and Iwama, 1972) (“ketamine spikes”; Fig. 2B, asterisk). The negative peaks of the ketamine spikes (280 ± 14 μV) were accompanied by similar spikes in the MEG (1.3 ± 0.18 pT) that could be visualized individually in the raw recording. In this figure, downward deflections of the MEG trace reflect a dorsal tangential magnetic field component pointed caudally. Thus, during the negative ECoG spike in the left hemisphere, the caudally pointed magnetic field was consistent with that generated by dipolar intradendritic currents perpendicular to and directed away from the cortical surface (Barth et al., 1986), producing clockwise magnetic fields (Fig. 2A, circle with arrows). The signal-to-noise ratios of ketamine spikes were 9.6 ± 1.3 and 76.2 ± 10.6 in the MEG and ECoG, respectively, when noise in the live animal (sampled between spikes) was used for comparison and 18.5 ± 2.4 and 197.2 ± 10.6, respectively, compared with noise after death. Examination of recordings before and after death indicated that higher noise in the live animal was due in part to spontaneous brain activity and not from other biological sources, such as respiration or cardiac artifact. Epileptiform spikes following cortical bicuculline methiodide injection at the cranial window were ∼3 times larger than ketamine spikes in the ECoG (Fig. 2C; 888 ± 32.7 μV; negative peak) with correspondingly large amplitude signals in the MEG (2.8 ± 0.18 pT). Signal-to-noise ratios in the MEG and ECoG of epileptiform spikes were also higher in the live (45.8 ± 1.4 and 606 ± 129.1, respectively) and killed (105.8 ± 20.7 and 855.6 ± 145.8, respectively) animals.
Simultaneous magnetic and electric recording of spontaneous spikes in the rat. A, Sagittal view of the rat skull (caudal is toward the right) showing positions of the 87Rb cell for MEG, and parietal cranial window for ECoG, recording in Experiment 1. The circle with arrows indicates the direction of magnetic fields expected during the negative peak of spikes in the left hemisphere ECoG. At the level of the MEG sensor, the tangential component of the magnetic field is pointed caudally. B, Top, Raw MEG (gray) and ECoG (black) of nonepileptiform ketamine spikes (*). Bottom, Five superimposed ketamine spikes (thin traces) and their average (thick traces). C, Similar to B but showing an example of a single epileptiform spike after focal injection of bicuculline methiodide via the cranial window. Top, Bottom, Single unaveraged spikes. D, Dorsal view showing bilateral placement of magnetrodes over the parietotemporal ridges and bilateral parietal ECoG electrodes. E, Averaged (n = 20) ketamine spikes using the left MEG response peak (arrow) to align the average. The MEG spike (gray) is lateralized to the left hemisphere and attenuated over the right. The averaged ECoG spike is confined to the left hemisphere. F, Similar to E but using the right MEG response peak (arrow) to align averaging. Note a complete shift of the MEG and ECoG response to the right hemisphere. The magnetic field is reversed to a rostral direction, as expected for a right hemisphere equivalent dipole reflecting intradendritic current flowing away from the cortical surface toward the deeper cortical layers.
Experiment 2 demonstrated the spatial resolution of μAM, with a capacity to discriminate small ketamine spikes originating in the opposite hemispheres. When we used the peaks of ∼20 MEG spikes recorded above the left hemisphere (Fig. 2E, arrow) to time-align averages of the bilateral ECoG and MEG, the electrical and magnetic signals indicated a large source in the left hemisphere (Fig. 2E, bottom), an attenuated response in the right MEG, and no response in the right ECoG (Fig. 2E, top). The converse was true when ketamine spikes in the right MEG sensor were used for averaging (Fig. 2F, arrow), resulting in a large ECoG response in the right hemisphere; MEG responses also lateralized largely to the right hemisphere, and no corresponding signal in the left hemisphere ECoG (Fig. 2F).
Discussion
The spatial resolution and sensitivity of the μAM demonstrated here are comparable with the most sensitive animal measurements of single epileptiform spikes performed with cryogenic technology (Nowak et al., 1999; Zwiener et al., 2000). Because the μAM operates without a cryogenic Dewar and can therefore be directly attached to the scalp as a magnetrode, its sensitivity for stationary recording of human epileptic discharge should also be comparable with present cryogenic MEG systems used in magnetically shielded environments. Yet, the small size of the μAM also makes it ideally suited to portable systems that could be worn for long-term nonstationary telemetry of brain pathology, such as epileptic seizures in humans. Recent advances in scalar optically pumped sensors have shown sensitivities <1 fT/Hz1/2 in centimeter-size cells (Sheng et al., 2013). The operation of scalar magnetometers improves the dynamic range and common-mode rejection of gradiometers. Microfabrication of such sensors can further reduce size and facilitate convenient recording in unshielded or minimally shielded environments required for practical human telemetry. Alternatively, further improvements in sensitivity of the current zero-field sensors to near the limit set by photon shot noise of 10 fT/Hz1/2 may also enable single-trial recordings of deeper brain structures in humans because the distance between sensor and scalp can be as small as 2.5 mm. Although these technical improvements will facilitate the use of atomic MEG for human telemetry, the present animal recordings justify and motivate this work. The μAM represents an advance in neuro-magnetometry with the potential to realize the originally anticipated application of MEG to the noninvasive presurgical evaluation of human epilepsy by localizing the source of seizures.
Footnotes
This work was supported by the U.S. Army Medical Research and Material Command Grant PR100040 and National Institute of Science and Technology. We thank John LeBlanc (Draper Laboratories) for the fabrication of sensor heads and Vishal Shah (QuSpin Inc.) for providing signal processing algorithms for the μAMs. This work is a contribution of National Institute of Science and Technology, an agency of the U.S. Government, and is not subject to copyright.