Skip to main content

Main menu

  • HOME
  • CONTENT
    • Early Release
    • Featured
    • Current Issue
    • Issue Archive
    • Collections
    • Podcast
  • ALERTS
  • FOR AUTHORS
    • Information for Authors
    • Fees
    • Journal Clubs
    • eLetters
    • Submit
  • EDITORIAL BOARD
  • ABOUT
    • Overview
    • Advertise
    • For the Media
    • Rights and Permissions
    • Privacy Policy
    • Feedback
  • SUBSCRIBE

User menu

  • Log in
  • My Cart

Search

  • Advanced search
Journal of Neuroscience
  • Log in
  • My Cart
Journal of Neuroscience

Advanced Search

Submit a Manuscript
  • HOME
  • CONTENT
    • Early Release
    • Featured
    • Current Issue
    • Issue Archive
    • Collections
    • Podcast
  • ALERTS
  • FOR AUTHORS
    • Information for Authors
    • Fees
    • Journal Clubs
    • eLetters
    • Submit
  • EDITORIAL BOARD
  • ABOUT
    • Overview
    • Advertise
    • For the Media
    • Rights and Permissions
    • Privacy Policy
    • Feedback
  • SUBSCRIBE
PreviousNext
TechSights

Wide. Fast. Deep: Recent Advances in Multiphoton Microscopy of In Vivo Neuronal Activity

Jérôme Lecoq, Natalia Orlova and Benjamin F. Grewe
Journal of Neuroscience 13 November 2019, 39 (46) 9042-9052; DOI: https://doi.org/10.1523/JNEUROSCI.1527-18.2019
Jérôme Lecoq
1Allen Institute for Brain Science, Seattle 98109, Washington,
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Jérôme Lecoq
Natalia Orlova
1Allen Institute for Brain Science, Seattle 98109, Washington,
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Natalia Orlova
Benjamin F. Grewe
2Institute of Neuroinformatics, UZH and ETH Zurich, Zurich 8057, Switzerland,
3Department of Electrical Engineering and Information Technology, ETH Zurich, Zurich 8092, Switzerland, and
4Faculty of Sciences, University of Zurich, Zurich 8057, Switzerland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Benjamin F. Grewe
  • Article
  • Figures & Data
  • Info & Metrics
  • eLetters
  • PDF
Loading

Abstract

Multiphoton microscopy (MPM) has emerged as one of the most powerful and widespread technologies to monitor the activity of neuronal networks in awake, behaving animals over long periods of time. MPM development spanned across decades and crucially depended on the concurrent improvement of calcium indicators that report neuronal activity as well as surgical protocols, head fixation approaches, and innovations in optics and microscopy technology. Here we review the last decade of MPM development and highlight how in vivo imaging has matured and diversified, making it now possible to concurrently monitor thousands of neurons across connected brain areas or, alternatively, small local networks with sampling rates in the kilohertz range. This review includes different laser scanning approaches, such as multibeam technologies as well as recent developments to image deeper into neuronal tissues using new, long-wavelength laser sources. As future development will critically depend on our ability to resolve and discriminate individual neuronal spikes, we will also describe a simple framework that allows performing quantitative comparisons between the reviewed MPM instruments. Finally, we provide our own opinion on how the most recent MPM developments can be leveraged at scale to enable the next generation of discoveries in brain function.

  • imaging
  • in vivo
  • multiphoton
  • two-photon

Introduction

When Adrian, Morison, and others performed the first in vivo recordings in anesthetized animals in the late 1930s, they were not able to record from more than a handful of neurons at a time (Adrian, 1936; Renshaw et al., 1940). Today, novel optical in vivo imaging methods enable routine high-resolution recordings of thousands of genetically identified individual neurons in the brain of awake, behaving animals over weeks and months (Trachtenberg et al., 2002; Holtmaat et al., 2009; Andermann et al., 2010; Attardo et al., 2015). Such capabilities illustrate the enormous advances that neurotechnologies have made over the last decades and how modern imaging tools open new avenues for neuroscience research. Compared with classical single-photon wide-field fluorescence microscopy, multiphoton microscopy (MPM) offers optical sectioning as well as depth penetration capabilities, two factors that greatly facilitate in vivo neural recordings deep in the intact living brain. Indeed, it was MPM that led the way for many important neuroscience discoveries revealing, among others, how new spines form during learning (Trachtenberg et al., 2002), how local spatial distribution of neuronal activity patterns represents sensory information (Ohki et al., 2005), and the role of other types of brain cells, such as microglia (Nimmerjahn et al., 2005).

Since the first demonstration of a working two-photon (2P) microscope in 1990 (Denk et al., 1990), MPM technology has made remarkable progress, including the development of large FOV (Sofroniew et al., 2016; Stirman et al., 2016) and multibeam optical designs (Lecoq et al., 2014; Voigt et al., 2015), as well as high-speed laser scanning (Duemani Reddy et al., 2008; Grewe et al., 2010) and spatiotemporal multiplexing approaches (Cheng et al., 2011) (Fig. 1). Miniaturized, head-mounted versions of 2P microscopes were also developed (Helmchen et al., 2001; Piyawattanametha et al., 2006) and have recently reemerged (Zong et al., 2017).

Figure 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 1.

Historical timeline highlighting major milestones of MPM development and its applications. Adapted with permission from Kaiser and Garrett (1961), Denk et al., (1990), Svoboda et al. (1997), Helmchen et al. (2001), Stosiek et al. (2003), Grewe et al. (2010), Sofroniew et al. (2016), Kobat et al. (2011), Chavarha et al. (2018).

Although standard 2P microscopes are widely used in neuroscience, MPM still has some drawbacks and limitations in terms of the temporal resolution, large-scale network monitoring, long-term applicability, and its ability to record neural activity during natural behaviors in freely moving animals. In this review, we will mainly focus on the technological advances made to the initial MPM microscopy setup (for a basic schematic, see Fig. 2) over the past 10 years that, in our opinion, have the potential to substantially improve neuroscience research in the near future. Our goal is not to convey all relevant technical details but rather to make neuroscience researchers aware of recent technological developments in MPM as well as their basic strengths and limitations. Therefore, we organized this review to focus on developments for recording from a larger number of neurons, recording neurons at higher speed, and recording deeper in the brain; we also provide a comparison of MPM techniques. Due to space limitations, we do not cover any mobile and/or miniaturized MPM approaches that are described previously (Zong et al., 2017).

Figure 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 2.

Schematic of a basic multiphoton microscope. An ultra-fast pulsed, near-infrared laser beam is directed onto a pair of X/Y scanning mirrors, which deflect the laser beam in two dimensions, depending on the voltage applied to their galvanometric elements. The beam then passes through the scan and tube lenses. The lens combination determines two fundamental parameters of scanning microscopy: the expansion of the beam entering the objective lens, which controls the size of the beam focus; and the scan angle of the laser beam into the objective, which is responsible for the size of FOV. Next, the laser beam is directed into the objective by a dichroic mirror that reflects near-infrared but passes light in the visible range. The objective lens focuses the short laser pulses into a tiny spot at the sample, resulting in multiphoton excitation of fluorescent indicator molecules. The emitted photons pass the dichroic mirror before being split into two channels (here, red and green) and are detected by two dedicated photomultiplier tubes (PMTs).

Meso-scale multiarea MPM approaches allow simultaneous in vivo recordings of thousands of neurons

Complex behavior often involves large brain networks that have both local and long-range connections and can span several millimeters in the mouse brain (Oh et al., 2014). To relate neuronal activity to behavior, it is crucial to simultaneously monitor the activity of such integrated long-range networks across many individual behavioral trials (Zhang et al., 2014). This experimental scientific need has fueled the development of new technologies that are capable of in vivo recordings from very large and distributed neuronal populations.

Interestingly, some of the biggest challenges in large-FOV optical systems were not optical. For example, only after the development of improved genetically encoded calcium indicators (GECIs) and related transgenic mouse lines (Dana et al., 2014; Madisen et al., 2015; Wekselblatt et al., 2016) did it become possible to label large populations of neuronal networks across multiple brain regions. In addition, early surgical protocols only provided limited optical access to a single isolated, superficial brain region either by thinning of the animal's skull (Stosiek et al., 2003) or by implanting a small cover glass as skull replacement (Trachtenberg et al., 2002). Such limitations were overcome by protocols using large optical windows (i.e., 5 mm diameter) implanted over a large portion of the mouse dorsal cortex (Goldey et al., 2014). Recently, a novel glass implant preparation known as the “crystal skull” allowed imaging of the entire dorsal cortex by implanting a single, curved glass window that mimicked the bone plate curvature (Kim et al., 2016). Thanks to these innovations, imaging of large populations of neurons is no longer limited by methods of sample preparation and cell labeling; however, we still lack the imaging technology to sample all these neurons efficiently. To address this bottleneck, several optical MPM design approaches have been developed over the last decade. Broadly, these fall into two categories: single-beam scanning techniques, which sequentially traverse large fields of neural tissue at high speed; and multibeam scanning techniques, which can simultaneously interrogate different locations of the neuronal tissue.

Single-beam MPM for large-scale interrogation of brain tissue

Initial attempts to scale up neuronal recordings focused on optimizing single-beam scanning and extending MPM to the third dimension. By axially moving the microscope objective with a fast piezo actuator, smart 3D line-scanning approaches (Göbel et al., 2007) allowed recording from ∼200 neurons at 10 Hz in the brain of live anesthetized rodents. An optimal use of the single laser-beam scanning time can be achieved by so-called random-access scanning (i.e., the laser beam is rapidly scanned over arbitrary selected points across the whole FOV). True random-access scanning can be realized with acousto-optic deflectors (AODs). These diffractive elements “deflect” a laser beam depending on the frequency of the sound wave generated by a piezo element that travels across a crystal positioned in the beam path. Two orthogonal AODs combine into a 2D scanner (Duemani Reddy et al., 2008; Grewe et al., 2010). If a pair of 2D-AODs are aligned in series with opposite sound propagation, 3D random-access imaging can be achieved with fast point-to-point transition times in the 10 μs range (Katona et al., 2012; Nadella et al., 2016). This allows spending the entire laser beam scan time on neuronal somata while avoiding scanning any unlabeled neuropil. However, all line- and point-scanning techniques are sensitive to sample motion, which is typically present when imaging in awake animals (Dombeck et al., 2007; Nimmerjahn et al., 2009). Various approaches were proposed to mitigate this issue, for example, by implementing a feedback loop for on-the-fly motion measurement and correction (Nadella et al., 2016) or by scanning local volumes around each area of interest (Szalay et al., 2016).

Despite these recent developments, fast raster scanning (i.e., sampling the FOV line by line) was favored for a long time, as motion artifacts could easily be resolved with custom postprocessing motion-correction algorithms (Kerr et al., 2005; Dombeck et al., 2007). This encouraged widespread adoption of galvanometric scanners operating at their resonance frequency, enabling fast, 2D scanning rates at ∼30 Hz (Fan et al., 1999). By combining such resonant scanners with an electrically tunable lens that shifts the axial focus of the laser beam by rapidly changing its shape, it became possible to scan two imaging planes at different depths at rates of up to 15 Hz, doubling the amount of recorded neurons (Grewe et al., 2011).

An alternative to inertia-limited axial scanners, such as piezo actuators or electrically tunable lenses, are spatial light modulators (SLMs) that are conjugated to the X/Y scanner and the Fourier plane of the objective (Liu et al., 2018). SLMs apply a variety of diffractive illumination patterns to the laser beam and thereby can shift MPM excitation laterally or axially. Contrary to inertia-limited electrically tunable lenses or piezo actuators, SLMs can provide fast (i.e., 5 ms) switching of the beam focus across the sample.

Another powerful concept for 3D imaging is remote focusing, which uses a second objective lens that focuses the beam on a mirror conjugated to the sample (Botcherby et al., 2012). Remote focusing not only corrects for optical aberrations introduced by the primary objective lens, but also provides a path for fast axial scanning as a lightweight mirror at the focus of the secondary objective can be moved by a fast actuator (e.g., a voice coil). Similarly, ultrasound lenses (Kong et al., 2015) refocus light by radially changing their refractive index. If used in a remote focusing setup, ultrasound lenses provide rapid axial scanning by changing the focus of the secondary objective.

An alternative approach to increase the number of recorded neurons is to enlarge the effective FOV by adjusting the microscope objective design. Commercially available objectives typically support scanning with a lateral extension of 500–800 × 500–800 μm2, but they do not allow large scanning angles without deteriorating the optical resolution of the whole imaging system (Stosiek et al., 2003; Ji et al., 2016). This issue was resolved through tailored in vivo 2P imaging objectives for very large FOVs (Tsai et al., 2015; Sofroniew et al., 2016; Stirman et al., 2016). Combined with more sophisticated single-beam scanner arrangements, 1000–2000 neurons were recorded within a 5 × 5 mm2 FOV (Sofroniew et al., 2016) at a sampling rate of 10 Hz (Fig. 3A,B). However, large-FOV objectives with sufficient NA require large and heavyweight optics that cannot be moved quickly for fast axial scanning. Therefore, next-generation large-FOV systems rely on remote focusing, SLMs, or electrically tunable lenses (Sofroniew et al., 2016; Stirman et al., 2016). To further extend the FOV, custom postobjective scanners, which position the scanning beam in the sample space using pairs of micro-mirrors moved by a rotating stage, were recently introduced (Terada et al., 2018).

Figure 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 3.

Large-area recording techniques. A, Schematic representation of a mesoscale (large FOV) imaging system known as 2P random-access microscope (2P-RAM). This imaging system supports slow scanning of a 5 mm2 FOV or high-resolution imaging of multiple 500 × 500 μm2 subregions with lateral separation of up to 4.5 mm. B, Low-magnification image acquired with the 2P-RAM system at 4.3 Hz frame rate, with an FOV of 5 mm2 (top). Four separate areas of 500 × 500 μm2 were sampled sequentially at 9.6 Hz (bottom). All data were acquired from a mouse expressing the calcium indicator GCaMP6f under a thy-1 promoter. C, Schematic of the dual optical axis MPM system. Two excitation arms are fully independent and have separate optical paths. To solve geometrical constraints, one of the objectives is coupled with micro-optical elements positioned onto the mouse's head. D, Two distal areas (S1 and M1) were imaged with the dual-axis imaging system in a GCaMP6s mouse. Scale bar, 300 μm. E, Schematic of a multibeam system with two simultaneous beams that scan two nearby areas. The scan areas were positioned within the FOV of a commercial objective and were laterally separated by <1 mm. F, Sequentially (top and middle) and simultaneously recorded images (bottom). Red boxes represent the area of FOV overlap. Computational algorithms and structural priors were used instead of excitation-emission encoding to separate the fluorescence signals of the two different planes. Scale bar, 50 μm. Adapted with permission from Lecoq et al. (2014), Sofroniew et al. (2016), and Yang et al. (2016).

Despite the possibility that new in vivo preparations, such as the crystal skull (Kim et al., 2016), would now even allow recording from ∼1 million cortical neurons, most of these single-beam scanning methods can only adequately detect calcium events across a very small subset of neurons simultaneously.

Multibeam MPM for sampling large neuronal population distributed across multiple distant sites

Perhaps one of the simplest multibeam imaging approaches is to use two independent pathways (Fig. 3C,D) by using micro-optics that provide the freedom of choosing two distant imaging sites (Lecoq et al., 2014). While this approach works well for distant recording sites (>1–2 mm apart), for adjacent areas, positioning multiple beams through a single objective can be more practical (Fig. 3E,F). For instance, with independent pathways, choosing new imaging areas requires a new optical alignment.

When designing multibeam scanning techniques, special attention must be given to crosstalk between the two excitation beams. Indeed, the presence of scattered photons in neural tissue (Oheim et al., 2001) limits the ability to later disambiguate multiple fluorescence emitters because fluorescence photons from one imaging plane are often detected in other imaging planes. This issue can be addressed either by post hoc source separation methods (Lecoq et al., 2014; Yang et al., 2016) or by temporal multiplexing schemes that support attributing the fluorescence signal to each individual scanning site (Cheng et al., 2011; Voigt et al., 2015). Temporal multiplexing can be easily implemented because MPM relies on high-power laser oscillators emitting very short (e.g., ∼100–200 femtosecond laser pulses at 80 MHz repetition rate). These short laser pulses excite a fluorescence signal that lasts a couple of nanoseconds (GFP fluorescence lifetime ≃ 3 ns) (Striker et al., 1999). This allows the use of multiple laser excitation beams simultaneously but requires the pulses of each beamlet to be delayed in time. This delay is crucial because it later allows temporal separation of the individual fluorescence signals that have been excited by the different beams (Cheng et al., 2011; Voigt et al., 2015; Stirman et al., 2016).

Recently, reverberation optical loops that circulate (loop) the excitation beams several times through a single delay line were introduced to efficiently distribute laser pulses that differ in time and axial position (Beaulieu et al., 2018). Such temporal approaches typically allow multiplexing of up to 4 beams and thus increase the number of imaged neurons by a factor of 4 accordingly. However, introducing >4 beams would lead to an increased overlap of fluorescence decays in time and thus limit the ability to distinguish the signal sources. Another major drawback that might have limited the broad adoption of multiplexing MPM is the need for custom-built high-bandwidth electronics operating at least twice the rate of the multiplexed laser pulse rate. When multiplexing 4 beams from a standard 80 MHz laser, this results in a sampling frequency of > 640 MHz to comply with Nyquist sampling criterion.

An alternative to online demultiplexing are offline computational algorithms that remove multiplexing crosstalk (Mukamel et al., 2009; Pnevmatikakis et al., 2016), which, in combination with SLM-based multibeam microscope, allow post hoc source separation (Yang et al., 2016) (Fig. 3E,F). In this case, all simultaneously acquired planes are overlapping during the acquisition, and calibration data need to be collected during the experiment to estimate a (de)mixing matrix.

Custom beam engineering to match the excitation beam to fluorescently labeled neuronal populations

When dealing with sparsely labeled tissue, raster scanning techniques mostly sample nonlabeled pixels. One possible approach to minimize this issue is to elongate the point-spread-function (PSF) in the axial direction to more efficiently sample the tissue along this imaging axis. This so-called “Bessel beam scanning” projects a volume of tissue into a single plane (Lu et al., 2017) and thus enables volumetric 400 × 400 × 100 μm3 MPM scan rates of 30–45 Hz. Although Bessel beams lack axial resolution along the z dimension, high-resolution image priors that are acquired with standard raster scanning can be used to retrieve the longitudinal coordinates of all recorded cells. Interestingly, Bessel beams at diameters well above the diffraction limit have also been successfully used in the context of densely labeled tissue. Indeed, diffraction-limited PSFs are often unnecessary for in vivo imaging, and a loss in resolution can be traded off for higher imaging speeds (Lecoq et al., 2014; Stirman et al., 2016). In the extreme case, one can fully abandon the diffraction-limited PSF and very efficiently scan the tissue with a large PSF (e.g., ∼5 × 5 × 10 μm3) that approximately matches the neuronal soma size. Recent works from the Vaziri group (Prevedel et al., 2016; Weisenburger et al., 2019) explored this direction. By combining temporal focusing, beam engineering, and smart laser pulse energy control with fast laser scanning, they could record neuronal activity from up to ∼12,000 neurons contained in a volume of 1 × 1 × 1.2 mm3. This combination of methods, however, is technically challenging because it requires temporal focusing and a custom laser source to maintain sufficient signal-to-noise ratio (SNR) levels. In addition, larger PSFs effectively harm the ability to distinguish the activity of nearby cells, which has to be accounted for.

Fast MPM approaches to resolve neuronal activity at the millisecond time scale

Neuronal networks in the brain process incoming stimuli within a few tens of milliseconds. To understand such fast neuronal dynamics, it is of key importance to investigate neuronal activity patterns with an adequate temporal resolution. With the introduction of fast calcium indicators (Chen et al., 2013; Dana et al., 2019) and recent improvements of voltage indicators (Gong et al., 2015; Lin and Schnitzer, 2016; Kannan et al., 2018), high-speed imaging has become a major direction in MPM development (St Pierre, 2017; Chavarha et al., 2018). Strategies to realize fast imaging speeds in vivo share principles similar to those used for imaging large FOVs, but instead use the extra recording bandwidth to increase the sampling rate. Fast MPM approaches can be categorized into three classes that include the following: (1) single-beam line scanning methods or AOD-based scanning that allow fast random-access scanning of neuronal populations; (2) multibeam scanning techniques that divide the FOV into subregions; and (3) fast tomography-based systems that image the sample from multiple angles.

Early MPM approaches to improve imaging speed aimed to maximize the time that the laser focus is scanning structures of interest. Increasing the imaging speed to kilohertz rates became possible with AOD scanners that allow the laser beam to move within microseconds and hence enable true high-speed random-access imaging (Iyer et al., 2006). For example, a pair of AODs allowed 2D in vivo population calcium imaging with near millisecond precision (Grewe et al., 2010). Moreover, 3D AOD-based systems with fast axial focusing could image populations of neurons at kilohertz rates (Fig. 4A–C) (Katona et al., 2012; Cotton et al., 2013; Nadella et al., 2016). For these reasons, AOD systems are particularly well suited for fast imaging of neuronal voltage signals. Recently, new genetically encoded voltage indicators have been combined with high-speed AOD-based MPM scanning approaches (Chavarha et al., 2018) to achieve fast population voltage imaging and to extract the spiking signals of individual neurons from the optical recordings with millisecond precision.

Figure 4.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 4.

Techniques for large-volume and high-speed scanning using AOD scanners and spatiotemporal multiplexing. A, Schematic representation of 3D AOD scanning used in random-access multiphoton (RAMP) microscopy. In AOD MPM systems, the number of neurons within a 3D volume directly determines the temporal sampling rates. B, MPM image overlaid with one of the scan patterns from A. C, Schematic of RAMP microscope. A commercial Ti:Al2O3 laser beam is expanded to fill the AOD aperture. All AODs (X1… Y2) are optically conjugated via 4f telescopes (T1–T4). A dichroic mirror separates the IR excitation and visible light detection paths. GaAsP photomultiplier tubes (PMTs) detect emitted photons (for more technical details, see original publication, Cotton et al., 2013). D, Principle of spatiotemporal multiplexing where a larger FOV is simultaneously scanned with 4 beams to increase data acquisition rate by a factor of 4. E, Simplified layout of the microscope with spatiotemporal multiplexing. The laser emits pulses at 80 MHz corresponding to a 12.5 ns interpulse interval. The initial beam is divided into 4 beams, which are delayed by ∼3 ns. All beams are recombined on a 2D-scanning module with different angular offset. In the imaging plane, this offset translates into a lateral displacement. Beam colors correspond to box colors in D. Emitted fluorescence is detected by PMTs, and photons are assigned to the correct image partition by a custom analog demultiplexing circuit based on their arrival time. For more details on temporal encoding and various techniques of multiplexing, see Cheng et al. (2011), Stirman et al. (2016), and Tsyboulski et al. (2018). Adapted with permission from Cheng et al. (2011) and Cotton et al. (2013).

The second most common approach to increase imaging speed is to divide the FOV into several subfields and to use multiple beams that simultaneously scan these (Fig. 4D,E). Such parallelized raster scanning has been demonstrated using a fast resonant scanning system with four temporally multiplexed beams, resulting in scan rates of up to 250 Hz over ∼400 μm2 FOV (Cheng et al., 2011). Theoretically, many more beams can be used to perform in vivo imaging, even at kHz rates (US Patent 9,820,652). However, such systems would require proportionally more laser power to maintain similar SNR levels for each individual beam, and thus might quickly exceed the heat-absorption-caused damage threshold for neuronal tissues (Podgorski and Ranganathan, 2016). To further increase the imaging bandwidth, more sophisticated multiplexing schemes have been proposed that either rely on a digital micromirror device that encodes a specific digital signature to each pixel by flipping individual micromirrors of the array (Ducros et al., 2013) or using interferometric AOD schemes that encode each beam with a different beat frequency (Tsyboulski et al., 2017). Although promising, these approaches have not yet been used extensively for in vivo recordings.

Another way to achieve kHz sampling rates relies on a tomography-like approach that scans the same probe from multiple angles (Kazemipour et al., 2019). These projections are acquired by four sequential line scans that traverse the FOV at different angles. This tomographic approach demonstrated dendritic imaging at rates >1 kHz across an FOV of 250 × 250 μm2 at a depth of up to 250 μm. However, one caveat of this approach is that the accuracy of signal reconstruction depends on the sparseness of the sample; for example, bright, clustered structures might not be reconstructed well. The tomography approach could be extended to larger FOVs; but as with other tomography methods, the number of reconstructed signal sources is limited to <1000 unless the number of angular projections is increased. Moreover, scanning with a complete line also implies delivering much higher average power to the sample, which might lead to increased thermal load as well as photobleaching and toxicity.

MPM approaches to probe deep cortical and sub-cortical brain areas in awake animals

Measuring complex activity patterns that underlie behavior often requires recording neuronal populations in deep cortical or subcortical brain areas. However, imaging deeper than ∼500 μm requires to adapt standard MPM techniques for an extended imaging depth. It is important to understand that the optical properties of tissue are the fundamental depth-limiting factors because both excitation and emission photons are highly scattered and absorbed by biological tissue. While the scattering problem could be addressed by increasing the laser intensity, this would introduce additional problems, such as sample heating, as well as out-of-focus and near-to-surface fluorescence volume excitation. Indeed, at least for 2P imaging, these issues seem most relevant for restricting the imaging depth (Theer and Denk, 2006), and the technologies that we discuss below attempt to resolve these limitations.

The most promising approach to increase the MPM imaging depth has been to shift the laser excitation light to longer wavelengths, for which scattering and absorption are substantially reduced (Lecoq and Schnitzer, 2011). Combined with the newest red-shifted GECIs, such as RCaMP, R-GECO, and jRCaMP (Akerboom et al., 2013; Dana et al., 2016), or related transgenic mice (Bethge et al., 2017), excitation wavelengths >1000 nm can be used to increase the 2P imaging depth down to ∼0.8 mm (Ouzounov et al., 2017). While early approaches (Kobat et al., 2011) used optical parametric oscillators that provided an excitation wavelength of 1280 nm to image vascular structures down to 1.6 mm deep, functional (calcium) imaging beyond a depth of 1 mm (e.g., in mouse CA1) requires even longer wavelengths between 1600 and 1800 nm that still lie within the water absorption minimum. In this regimen, 3-photon (3P) absorption will become the major process of fluorescence excitation. Indeed, out-of-focus and near-surface fluorescence volume excitation, the two biggest depth-limiting factors for 2P imaging, are substantially reduced for 3P imaging (Theer and Denk, 2006). Further increasing the laser excitation wavelength to 1700 nm can be achieved by another laser type, a photonic-crystal pumped by an erbium fiber laser. Using this approach, Horton et al. (2013) demonstrated 3P in vivo calcium imaging of neuronal and vascular structures up a depth of 1.4 mm (Fig. 5). However, 3P imaging also requires an order-of-magnitude higher pulse energies to achieve the same fluorescence signal as 2P excitation. This is because fluorophore absorption cross-sections are substantially smaller for 3P excitation (compare 3P ∼10−83 cm6 s2/photon2 vs 2P ∼10−49 cm4 s/photon for GFP) (Cheng et al., 2014). Such high pulse energies can be achieved through optical parametric amplifiers that provide pulse energies of hundreds of nanojoules within a tunable range of 1000–1600 nm. While the use of optical parametric amplifiers enabled structural in vivo imaging up to a depth of 1.5 mm at 1215 nm wavelength (Miller et al., 2017), more recently functional 3P imaging of mouse hippocampal and deep cortical areas at a depth of 0.7–1.2 mm has been demonstrated (Ouzounov et al., 2017; Yildirim et al., 2019) (Fig. 5).

Figure 5.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 5.

Advanced 3P microscopy approaches to record neuronal activity deep in the mouse brain. A, Schematic of the 3P deep-tissue imaging approach. B, Direct comparison of the signal-to-background ratio for 2P and 3P imaging of GCaMP6s-labeled neurons 780 μm deep in mouse cortex. “Background” arrow indicates the out-of-focus fluorescence measured inside a blood vessel. “Background & Neuropil” arrow indicates neuropil and the out-of-focus background. The two images have a comparable number of signal photons and are displayed with the same contrast settings (top 0.4% saturation). Scale bars, 30 μm. C, 3P z stack acquired with 2 μm steps. Green represents GCaMP6s signal. Magenta represents the label-free third harmonic generation signal from the blood vessels in the visual cortex and myelin fibers in the white matter. Scale bar, 100 μm. D, Selection of 2D images from various layers (L2/3, L4, L5, and L6). Scale bar, 50 μm. FOV in all lateral images is 250 μm. Adapted with permission from Yildirim et al. (2019), and Ouzounov et al. (2017).

To summarize, functional 3P microscopy is substantially more demanding than its structural counterpart as it requires fast scanning and short pixel dwell times to adequately sample the neuronal activity in time. In addition, pulse energies must be significantly higher (5–10 nJ compared with 0.3–3 nJ for standard 2P imaging) to gather enough signal during each pixel dwell time. Such high pulse energies also seem to increase photodamage in a nonlinear manner, which is more likely an issue in 3P imaging than heat absorption (Podgorski and Ranganathan, 2016; Yildirim et al., 2019). Another drawback of using optical parametric amplifiers for 3P imaging is that they can lower the pulse repetition rate to ∼400–800 kHz, which limits the number of pixels that can be scanned across the FOV within a certain time interval, thereby effectively reducing the FOV size. However, newer 3P optical parametric amplifiers promise pulse repetition rates of up to 5 MHz. As for other technical developments, a combination of the techniques described above is currently explored to circumvent these limitations. For instance, combining Bessel beam with 3P absorption offers higher volume rate (Chen et al., 2018; Rodríguez et al., 2018).

Comparison of MPM techniques

The core idea of this review is to provide an overview of the most recent MPM developments and applications with the goal to enable researchers to select the best suitable method for their in vivo imaging experiments. However, a direct method comparison is difficult because comparative standards have not yet been established in the field. First, individual MPM approaches cannot be directly compared because experimental conditions vary across studies (e.g., fluorescent indicators, wavelengths, imaging depths, microscope objectives, photomultiplier tubes, and preamplifiers). Second, in addition to the sampling rate and number of neurons, key metrics needed to evaluate the quality of in vivo recordings, such as the ability to discriminate neuronal activity signals from background, are often not reported. Third, typically only 1 or 2 performance data points (sampling rate and number of neurons) are reported per study, and the extrapolation to alternative MPM technologies proves difficult. Although a direct method comparison is hardly possible, it is clear that most recent MPM developments aimed to optimize the following three parameters: (1) the number of neurons that can be recorded; (2) the sampling rate and temporal precision to resolve neuronal activity; and (3) the reliability of detecting calcium events from the fluorescence signals and extracting individual neuronal spikes. While (1) and (2) are straightforward to evaluate, (3) is frequently underappreciated due to the lack of data. In the following section, we will attempt to analytically describe a framework to improve the standardization of reporting (3), the signal detection performance.

While simple metrics, such as the single action potential (1AP) SNR, were initially used to describe the ability to detect neuronal signals from fluorescence traces, recent modeling work suggests the use of more sophisticated metrics, which are commonly known in signal detection theory as the sensitivity index, or d′ (Wilt et al., 2013). d′ quantifies the ability to distinguish individual neuronal calcium events from baseline noise. Under certain assumptions (e.g., normal distribution), d′ can be expressed as Embedded Image, where μ and σ are the mean and the SD for the 1AP and the baseline signal distributions. Since σ is mainly driven by shot noise, one can assume σN = σS, which simplifies the equation to Embedded Image. As a rule of thumb, a d′ of 5 relates to a correct 1AP detection rate of ∼99%.

If the sampling rate f is substantially higher than the decay time constant τind of the calcium indicator (f × τind ≫ 1), d′ can be approximated as Embedded Image (Wilt et al., 2013). Here, F0 is the somatic background fluorescence in photons/s and ΔF/F is the peak fluorescence change detected for a single AP. This means that, to achieve a d′ of 5 with GCaMP6f (1AP ΔF/F ∼15%, τind ∼0.2 s), the laser intensity or signal integration times need to be adjusted so that at least F0 >11,000 photons/s/cell are detected under baseline conditions. If the main interest is to reliably detect short bursts of, for example, 5 APs with the same d′, at least F0 > 2200 photons/s/cell are required assuming a linear signal summation. In the following paragraph, we use standardized data acquired at the Allen Brain Observatory (de Vries et al., 2018) to provide an idea of how to calculate typical F0 numbers and associated theoretical d′ values for standard MPM systems.

The standard MPM systems at the Allen Institute for Brain Science use resonant galvanometric scanners, capable of scanning a single excitation laser beam across a 400 × 400 μm2 FOV (512 × 512 pixels) at 30 Hz (de Vries et al., 2018). Taking into account the gain and offset of the data acquisition (photomultiplier tubes and preamplifiers), we calculated the median photon flux per pixel directly from the motion-corrected fluorescence data (for code to apply this method to any MPM data, see https://github.com/AllenInstitute/QC_2P) to be, on average, 0.89 photons per pixel per dwell time (n = 2080 experiments across both excitatory and inhibitory transgenic cre lines expressing GCaMP6f in all layers of the mouse visual cortex). In this case, a cellular soma spans ∼100 pixels2 that provide a F0 ≅ 2670 photons/s/cell or a theoretical d′ value of 2.4, which is well within a range where ≥5 APs are reliably detected (d′ ∼ 5).

A simple way to increase d′ is to decrease the FOV size from 500 × 500 μm2 to 180 × 180 μm2, which would result in a 5× increase of the number of pixels per neuron and thus a 5× higher number of detected photons. This would result in F0 > 13,000 photons/s/cell and thus a theoretical d′ of 5.4. However, shrinking the FOV will also decrease the number of recorded neurons by a factor of 5. Alternatively, slower versions of GCaMP6 (i.e., GCaMP6s) can be used, providing an approximate 1AP ΔF/F change of 25% and a τind ∼ 0.6 s (Chen et al., 2013). Because of the larger ΔF/F and τind, GCaMP6s can achieve a d′ of 5 with 1.7 times less recorded pixels per cell compared with GCaMP6f. However, using slower GECIs comes at the cost of a reduced temporal precision of detecting neuronal signals.

The best temporal precision can be achieved when using genetically encoded voltage indicators that directly report fast changes of the neuronal membrane voltage. For example, using an improved voltage indicator, ASAP3, in vivo population voltage imaging of 3–5 neurons was recently demonstrated at a rate of ∼3 kHz, which is ∼8 times faster than the temporal dynamics of the recorded voltage transients (Chavarha et al., 2018). The approximation introduced above (Wilt et al., 2013) will only hold if f × τind ≫ 1. Thus, if lower sampling frequencies are used, even higher photon counts are required to achieve a d′ of 5. However, even when imaging at 3 kHz, imaging with a d′ value of 5 can still falsely detect ∼10 spikes per second, since each imaging frame can yield false positives. Thus, for fast voltage imaging, higher d′ values are required. For example, sampling at 3 kHz, Chavarha et al. (2018) reported a d′ of 9, which independent of the actual neuronal activity resulted in a false positive spike detection rate of one spike every 200 s and a false negative detection rate of 1.5 × 10−6 spikes per second. Given the reported 1AP ΔF/F of 9% and a decay time of ∼4 ms, the number of baseline photons that should be collected are F0 > 5000 photons/ms/cell. These simple calculations illustrate that voltage imaging of large populations (>50 neurons) will require substantially more advanced MPM solutions or alternative imaging approaches that provide at least 1–2 orders of magnitude better imaging capabilities than current high-speed MPM approaches.

We summarize the properties of recent MPM developments in Table 1 and Figure 6, which shows that the number of recorded neurons is generally inversely related to the imaging rate. In this figure, we only report features and numbers that have been published or, in some cases, that we obtained by contacting the authors to clarify their method. Our comparison of the different microscopy systems is not optimal because the reported data have been acquired under different imaging conditions (e.g., differently labeling techniques, expression levels, and approaches to quantify and report SNR). The difficulty in comparing these methods also underlines the increasing need to develop a set of standardized evaluation metrics to report advances in MPM.

View this table:
  • View inline
  • View popup
Table 1.

Representative publications for multiple recent advanced MPM techniquesa

Figure 6.
  • Download figure
  • Open in new tab
  • Download powerpoint
Figure 6.

Comparison of MPM techniques in terms of number of neurons and recording speed. Mapping of recently developed MPM systems onto a 2D space (number of imaged neurons vs sampling rate). For some cases, we converted the imaging FOV into number of neurons assuming a densely labeled GCaMP mouse line, such as Slc17a7:Ai148 where, on average, 200 labeled neurons are detected in a 500 × 500 μm2 FOV. Conventional galvo-galvo and resonant-galvo systems are included to emphasize the progress in technology development over the last decade. With the increasing number of scanning beams, the plot shifts toward higher speed or number of neurons, depending on the microscope configuration. Square markers represent published data. Clouds represent hypothetical interpolation between the reported data points.

Discussion

The future of large-scale neuronal recording most certainly rests upon a combination of the above-mentioned techniques. Newly designed objectives lenses offer an extended view into neuronal tissues that can only be leveraged with high-bandwidth recording techniques. SLMs are promising tools providing flexibility to create multiple beams and even allow us to look deeper into tissue using concepts of adaptive optics (Ji et al., 2010; Liu et al., 2018). Similarly, multiplexing techniques combined with sophisticated scanning technologies, large FOV objective lens designs, and excitation beam shaping offer the opportunity to further increase the number of simultaneously recorded neurons to many thousands of neurons (Tsyboulski et al., 2018; Weisenburger et al., 2019). It is reasonable to expect that, in mice, entire cortical areas could be recorded in the very near future. As the sophistication of our instruments increases, the need for comparative quantitative standards becomes central to the development of future technologies. With integrated optical and mechanical computer-aided design tools, such as Zemax, Solidworks (Dassault Systèmes), and Creo (Parametric Technology), it becomes incredibly important to predict all performance aspects of new instruments before substantial financial investments are made.

Finally, given today's remarkable technological and experimental possibilities, the next generation of MPM instruments should be driven by neurobiological questions arising in the field. Indeed, multiphoton microscopes have matured from their initial design to showcase different, complementary capabilities. Instruments to image vast populations of neurons are designed very differently than those that aim to image dendritic processes.

Given the increased complexity of MPM instruments, their limited availability should not prevent us from concentrating on important neuroscience questions. In this regard, an academic observatory model currently used in high-energy physics and astronomy offers a unique opportunity to gain access to complex instrumentation and experimental techniques while encouraging data standardization. The idea of a brain observatory model is to open access to advanced experimentation based on unbiased scientific reviews of submitted proposals. Supportive staff can then perform the experiments while meeting high-quality standards and standardized operating procedures. Such an observatory approach is currently explored for systems neuroscience at the Allen Institute for Brain Science through the OpenScope project.

While monitoring the activity of thousands of neurons chronically was a distant vision just a few decades ago, it is now routinely performed in many laboratories through MPM of genetically encoded calcium reporters of neuronal activity. Significant technological developments have recently broadened our access to neuronal tissues and greatly increased the number of different techniques available. These allow for sampling larger populations of neurons, accessing neuronal networks located deep in the neuronal tissue, or achieving faster recording rates. Future developments will likely combine these techniques to further broaden our access to neuronal activity. Tight integration of these developments with neuroscience research is bound to transform our understanding of brain function.

Footnotes

  • We thank Kevin Takasaki and Peter Saggau (Allen Institute for Brain Science) for providing helpful comments on the manuscript; and Bénédicte Rossi for providing scientific illustrations. The authors thank the Allen Institute for Brain Science founders, Paul G. Allen and Jody Allen, for their vision, encouragement, and support.

  • The authors declare no competing financial interests.

  • Correspondence should be addressed to Jérôme Lecoq at jeromel{at}alleninstitute.org

References

  1. ↵
    1. Adrian ED
    (1936) The spread of activity in the cerebral cortex. J Physiol 88:127–161. doi:10.1113/jphysiol.1936.sp003427 pmid:16994811
    OpenUrlCrossRefPubMed
  2. ↵
    1. Akerboom J,
    2. Carreras Calderón N,
    3. Tian L,
    4. Wabnig S,
    5. Prigge M,
    6. Tolö J,
    7. Gordus A,
    8. Orger MB,
    9. Severi KE,
    10. Macklin JJ,
    11. Patel R,
    12. Pulver SR,
    13. Wardill TJ,
    14. Fischer E,
    15. Schüler C,
    16. Chen TW,
    17. Sarkisyan KS,
    18. Marvin JS,
    19. Bargmann CI,
    20. Kim DS, et al
    . (2013) Genetically encoded calcium indicators for multi-color neural activity imaging and combination with optogenetics. Front Mol Neurosci 6:2. doi:10.3389/fnmol.2013.00002 pmid:23459413
    OpenUrlCrossRefPubMed
  3. ↵
    1. Andermann ML,
    2. Kerlin AM,
    3. Reid RC
    (2010) Chronic cellular imaging of mouse visual cortex during operant behavior and passive viewing. Front Cell Neurosci 4:3. doi:10.3389/fncel.2010.00003 pmid:20407583
    OpenUrlCrossRefPubMed
  4. ↵
    1. Attardo A,
    2. Fitzgerald JE,
    3. Schnitzer MJ
    (2015) Impermanence of dendritic spines in live adult CA1 hippocampus. Nature 523:592–596. doi:10.1038/nature14467 pmid:26098371
    OpenUrlCrossRefPubMed
  5. ↵
    1. Beaulieu DR,
    2. Davison IG,
    3. Bifano TG,
    4. Mertz J
    (2018) Simultaneous multiplane imaging with reverberation multiphoton microscopy. arxiv. Available at https://arxiv.org/abs/1812.05162.
  6. ↵
    1. Bethge P,
    2. Carta S,
    3. Lorenzo DA,
    4. Egolf L,
    5. Goniotaki D,
    6. Madisen L,
    7. Voigt FF,
    8. Chen JL,
    9. Schneider B,
    10. Ohkura M,
    11. Nakai J,
    12. Zeng H,
    13. Aguzzi A,
    14. Helmchen F
    (2017) An R-CaMP1.07 reporter mouse for cell-type-specific expression of a sensitive red fluorescent calcium indicator. PLoS One 12:e0179460. doi:10.1371/journal.pone.0179460 pmid:28640817
    OpenUrlCrossRefPubMed
  7. ↵
    1. Botcherby EJ,
    2. Smith CW,
    3. Kohl MM,
    4. Débarre D,
    5. Booth MJ,
    6. Juskaitis R,
    7. Paulsen O,
    8. Wilson T
    (2012) Aberration-free three-dimensional multiphoton imaging of neuronal activity at kHz rates. Proc Natl Acad Sci U S A 109:2919–2924. doi:10.1073/pnas.1111662109 pmid:22315405
    OpenUrlAbstract/FREE Full Text
  8. ↵
    1. Chavarha M,
    2. Villette V,
    3. Dimov I,
    4. Pradhan L,
    5. Evans S,
    6. Shi D,
    7. Yang R,
    8. Chamberland S,
    9. Bradley J,
    10. Mathieu B,
    11. St-Pierre F,
    12. Schnitzer MJ,
    13. Bi G,
    14. Toth K,
    15. Ding J,
    16. Dieudonne S,
    17. Lin M
    (2018) Fast two-photon volumetric imaging of an improved voltage indicator reveals electrical activity in deeply located neurons in the awake brain. bioRxiv. Advance online publication. Retrieved October 17, 2018. doi: 10.1101/445064. doi:10.1101/445064
    OpenUrlAbstract/FREE Full Text
  9. ↵
    1. Chen B,
    2. Huang X,
    3. Gou D,
    4. Zeng J,
    5. Chen G,
    6. Pang M,
    7. Hu Y,
    8. Zhao Z,
    9. Zhang Y,
    10. Zhou Z,
    11. Wu H,
    12. Cheng H,
    13. Zhang Z,
    14. Xu C,
    15. Li Y,
    16. Chen L,
    17. Wang A
    (2018) Rapid volumetric imaging with Bessel-beam three-photon microscopy. Biomed Opt Express 9:1992–2000. doi:10.1364/BOE.9.001992 pmid:29675334
    OpenUrlCrossRefPubMed
  10. ↵
    1. Chen TW,
    2. Wardill TJ,
    3. Sun Y,
    4. Pulver SR,
    5. Renninger SL,
    6. Baohan A,
    7. Schreiter ER,
    8. Kerr RA,
    9. Orger MB,
    10. Jayaraman V,
    11. Looger LL,
    12. Svoboda K,
    13. Kim DS
    (2013) Ultrasensitive fluorescent proteins for imaging neuronal activity. Nature 499:295–300. doi:10.1038/nature12354 pmid:23868258
    OpenUrlCrossRefPubMed
  11. ↵
    1. Cheng A,
    2. Tiago Gonçalves JT,
    3. Golshani P,
    4. Arisaka K,
    5. Portera-Cailliau C
    (2011) Simultaneous two-photon calcium imaging at different depths with spatiotemporal multiplexing. Nat Methods 8:139–142. doi:10.1038/nmeth.1552 pmid:21217749
    OpenUrlCrossRefPubMed
  12. ↵
    1. Cheng LC,
    2. Horton NG,
    3. Wang K,
    4. Chen SJ,
    5. Xu C
    (2014) Measurements of multiphoton action cross sections for multiphoton microscopy. Biomed Opt Express 5:3427–3433. doi:10.1364/BOE.5.003427 pmid:25360361
    OpenUrlCrossRefPubMed
  13. ↵
    1. Cotton RJ,
    2. Froudarakis E,
    3. Storer P,
    4. Saggau P,
    5. Tolias AS
    (2013) Three-dimensional mapping of microcircuit correlation structure. Front Neural Circuits 7:151. doi:10.3389/fncir.2013.00151 pmid:24133414
    OpenUrlCrossRefPubMed
  14. ↵
    1. Dana H,
    2. Chen TW,
    3. Hu A,
    4. Shields BC,
    5. Guo C,
    6. Looger LL,
    7. Kim DS,
    8. Svoboda K
    (2014) Thy1-GCaMP6 transgenic mice for neuronal population imaging in vivo. PLoS One 9:e108697. doi:10.1371/journal.pone.0108697 pmid:25250714
    OpenUrlCrossRefPubMed
  15. ↵
    1. Dana H,
    2. Mohar B,
    3. Sun Y,
    4. Narayan S,
    5. Gordus A,
    6. Hasseman JP,
    7. Tsegaye G,
    8. Holt GT,
    9. Hu A,
    10. Walpita D,
    11. Patel R,
    12. Macklin JJ,
    13. Bargmann CI,
    14. Ahrens MB,
    15. Schreiter ER,
    16. Jayaraman V,
    17. Looger LL,
    18. Svoboda K,
    19. Kim DS
    (2016) Sensitive red protein calcium indicators for imaging neural activity. Elife 5:312727. doi:10.7554/eLife.12727 pmid:27011354
    OpenUrlCrossRefPubMed
  16. ↵
    1. Dana H,
    2. Sun Y,
    3. Mohar B,
    4. Hulse BK,
    5. Kerlin AM,
    6. Hasseman JP,
    7. Tsegaye G,
    8. Tsang A,
    9. Wong A,
    10. Patel R,
    11. Macklin JJ,
    12. Chen Y,
    13. Konnerth A,
    14. Jayaraman V,
    15. Looger LL,
    16. Schreiter ER,
    17. Svoboda K,
    18. Kim DS
    (2019) High-performance GFP-based calcium indicators for imaging activity in neuronal populations and microcompartments. Nat Methods 16:649–657. doi:10.1038/s41592-019-0435-6 pmid:31209382
    OpenUrlCrossRefPubMed
  17. ↵
    1. de Vries SEJ,
    2. Lecoq J,
    3. Buice MA,
    4. Groblewski PA,
    5. Ocker GK,
    6. Oliver M,
    7. Feng D,
    8. Cain N,
    9. Ledochowitsch P,
    10. Millman D,
    11. Roll K,
    12. Garrett M,
    13. Keenan T,
    14. Kuan L,
    15. Mihalas S,
    16. Olsen S,
    17. Thompson C,
    18. Wakeman W,
    19. Waters J,
    20. Williams D, et al
    . (2018) A large-scale, standardized physiological survey reveals higher order coding throughout the mouse visual cortex. bioRxiv. Available at https://europepmc.org/abstract/PPR/PPR7988.
  18. ↵
    1. Denk W,
    2. Strickler JH,
    3. Webb WW
    (1990) Two-photon laser scanning fluorescence microscopy. Science 248:73–76. doi:10.1126/science.2321027 pmid:2321027
    OpenUrlAbstract/FREE Full Text
  19. ↵
    1. Dombeck DA,
    2. Khabbaz AN,
    3. Collman F,
    4. Adelman TL,
    5. Tank DW
    (2007) Imaging large-scale neural activity with cellular resolution in awake, mobile mice. Neuron 56:43–57. doi:10.1016/j.neuron.2007.08.003 pmid:17920014
    OpenUrlCrossRefPubMed
  20. ↵
    1. Ducros M,
    2. Goulam Houssen Y,
    3. Bradley J,
    4. de Sars V,
    5. Charpak S
    (2013) Encoded multisite two-photon microscopy. Proc Natl Acad Sci U S A 110:13138–13143. doi:10.1073/pnas.1307818110 pmid:23798397
    OpenUrlAbstract/FREE Full Text
  21. ↵
    1. Duemani Reddy G,
    2. Kelleher K,
    3. Fink R,
    4. Saggau P
    (2008) Three-dimensional random access multiphoton microscopy for functional imaging of neuronal activity. Nat Neurosci 11:713–720. doi:10.1038/nn.2116 pmid:18432198
    OpenUrlCrossRefPubMed
  22. ↵
    1. Fan GY,
    2. Fujisaki H,
    3. Miyawaki A,
    4. Tsay RK,
    5. Tsien RY,
    6. Ellisman MH
    (1999) Video-rate scanning two-photon excitation fluorescence microscopy and ratio imaging with chameleons. Biophys J 76:2412–2420. doi:10.1016/S0006-3495(99)77396-0 pmid:10233058
    OpenUrlCrossRefPubMed
  23. ↵
    1. Göbel W,
    2. Kampa BM,
    3. Helmchen F
    (2007) Imaging cellular network dynamics in three dimensions using fast 3D laser scanning. Nat Methods 4:73–79. doi:10.1038/nmeth989 pmid:17143280
    OpenUrlCrossRefPubMed
  24. ↵
    1. Goldey GJ,
    2. Roumis DK,
    3. Glickfeld LL,
    4. Kerlin AM,
    5. Reid RC,
    6. Bonin V,
    7. Schafer DP,
    8. Andermann ML
    (2014) Removable cranial windows for long-term imaging in awake mice. Nat Protoc 9:2515–2538. doi:10.1038/nprot.2014.165 pmid:25275789
    OpenUrlCrossRefPubMed
  25. ↵
    1. Gong Y,
    2. Huang C,
    3. Li JZ,
    4. Grewe BF,
    5. Zhang Y,
    6. Eismann S,
    7. Schnitzer MJ
    (2015) High-speed recording of neural spikes in awake mice and flies with a fluorescent voltage sensor. Science 350:1361–1366. doi:10.1126/science.aab0810 pmid:26586188
    OpenUrlAbstract/FREE Full Text
  26. ↵
    1. Grewe BF,
    2. Langer D,
    3. Kasper H,
    4. Kampa BM,
    5. Helmchen F
    (2010) High-speed in vivo calcium imaging reveals neuronal network activity with near-millisecond precision. Nat Methods 7:399–405. doi:10.1038/nmeth.1453 pmid:20400966
    OpenUrlCrossRefPubMed
  27. ↵
    1. Grewe BF,
    2. Voigt FF,
    3. van 't Hoff M,
    4. Helmchen F
    (2011) Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens. Biomed Opt Express 2:2035–2046. doi:10.1364/BOE.2.002035 pmid:21750778
    OpenUrlCrossRefPubMed
  28. ↵
    1. Helmchen F,
    2. Fee MS,
    3. Tank DW,
    4. Denk W
    (2001) A miniature head-mounted two-photon microscope: high-resolution brain imaging in freely moving animals. Neuron 31:903–912. doi:10.1016/S0896-6273(01)00421-4 pmid:11580892
    OpenUrlCrossRefPubMed
  29. ↵
    1. Holtmaat A,
    2. Bonhoeffer T,
    3. Chow DK,
    4. Chuckowree J,
    5. De Paola V,
    6. Hofer SB,
    7. Hübener M,
    8. Keck T,
    9. Knott G,
    10. Lee WC,
    11. Mostany R,
    12. Mrsic-Flogel TD,
    13. Nedivi E,
    14. Portera-Cailliau C,
    15. Svoboda K,
    16. Trachtenberg JT,
    17. Wilbrecht L
    (2009) Long-term, high-resolution imaging in the mouse neocortex through a chronic cranial window. Nat Protoc 4:1128–1144. doi:10.1038/nprot.2009.89 pmid:19617885
    OpenUrlCrossRefPubMed
    1. Horton NG,
    2. Kobat D,
    3. Wang K,
    4. Xu C
    (2012) In vivo, deep tissue three-photon imaging at the 1700-nm spectral window. In: Biomedical optics and 3-D imaging. Washington, DC: OSA.
  30. ↵
    1. Horton NG,
    2. Wang K,
    3. Kobat D,
    4. Clark CG,
    5. Wise FW,
    6. Schaffer CB,
    7. Xu C
    (2013) In vivo three-photon microscopy of subcortical structures within an intact mouse brain. Nat Photonics 7:205–209. doi:10.1038/nphoton.2012.336 pmid:24353743
    OpenUrlCrossRefPubMed
  31. ↵
    1. Iyer V,
    2. Hoogland TM,
    3. Saggau P
    (2006) Fast functional imaging of single neurons using random-access multiphoton (RAMP) microscopy. J Neurophysiol 95:535–545. doi:10.1152/jn.00865.2005 pmid:16221746
    OpenUrlCrossRefPubMed
  32. ↵
    1. Ji N,
    2. Milkie DE,
    3. Betzig E
    (2010) Adaptive optics via pupil segmentation for high-resolution imaging in biological tissues. Nat Methods 7:141–147. doi:10.1038/nmeth.1411 pmid:20037592
    OpenUrlCrossRefPubMed
  33. ↵
    1. Ji N,
    2. Freeman J,
    3. Smith SL
    (2016) Technologies for imaging neural activity in large volumes. Nat Neurosci 19:1154–1164. doi:10.1038/nn.4358 pmid:27571194
    OpenUrlCrossRefPubMed
  34. ↵
    1. Kaiser W,
    2. Garrett CGB
    (1961) Two-photon excitation in Ca F 2: Eu 2+. Physical review letters 76: 229
    OpenUrl
  35. ↵
    1. Kannan M,
    2. Vasan G,
    3. Huang C,
    4. Haziza S,
    5. Li JZ,
    6. Inan H,
    7. Schnitzer MJ,
    8. Pieribone VA
    (2018) Fast, in vivo voltage imaging using a red fluorescent indicator. Nat Methods 15:1108–1116. doi:10.1038/s41592-018-0188-7 pmid:30420685
    OpenUrlCrossRefPubMed
  36. ↵
    1. Katona G,
    2. Szalay G,
    3. Maák P,
    4. Kaszás A,
    5. Veress M,
    6. Hillier D,
    7. Chiovini B,
    8. Vizi ES,
    9. Roska B,
    10. Rózsa B
    (2012) Fast two-photon in vivo imaging with three-dimensional random-access scanning in large tissue volumes. Nat Methods 9:201–208. doi:10.1038/nmeth.1851 pmid:22231641
    OpenUrlCrossRefPubMed
  37. ↵
    1. Kazemipour A,
    2. Novak O,
    3. Flickinger D,
    4. Marvin JS,
    5. Abdelfattah AS,
    6. King J,
    7. Borden PM,
    8. Kim JJ,
    9. Al-Abdullatif SH,
    10. Deal PE,
    11. Miller EW,
    12. Schreiter ER,
    13. Druckmann S,
    14. Svoboda K,
    15. Looger LL,
    16. Podgorski K
    (2019) Kilohertz frame-rate two-photon tomography. Nat Methods 16:778–786. doi:10.1038/s41592-019-0493-9 pmid:31363222
    OpenUrlCrossRefPubMed
  38. ↵
    1. Kerr JN,
    2. Greenberg D,
    3. Helmchen F
    (2005) Imaging input and output of neocortical networks in vivo. Proc Natl Acad Sci U S A 102:14063–14068. doi:10.1073/pnas.0506029102 pmid:16157876
    OpenUrlAbstract/FREE Full Text
  39. ↵
    1. Kim TH,
    2. Zhang Y,
    3. Lecoq J,
    4. Jung JC,
    5. Li J,
    6. Zeng H,
    7. Niell CM,
    8. Schnitzer MJ
    (2016) Long-term optical access to an estimated one million neurons in the live mouse cortex. Cell Rep 17:3385–3394. doi:10.1016/j.celrep.2016.12.004 pmid:28009304
    OpenUrlCrossRefPubMed
  40. ↵
    1. Kobat D,
    2. Horton NG,
    3. Xu C
    (2011) In vivo two-photon microscopy to 1.6-mm depth in mouse cortex. J Biomed Opt 16:106014. doi:10.1117/1.3646209 pmid:22029361
    OpenUrlCrossRefPubMed
  41. ↵
    1. Kong L,
    2. Tang J,
    3. Little JP,
    4. Yu Y,
    5. Lämmermann T,
    6. Lin CP,
    7. Germain RN,
    8. Cui M
    (2015) Continuous volumetric imaging via an optical phase-locked ultrasound lens. Nat Methods 12:759–762. doi:10.1038/nmeth.3476 pmid:26167641
    OpenUrlCrossRefPubMed
  42. ↵
    1. Lecoq J,
    2. Schnitzer MJ
    (2011) An infrared fluorescent protein for deeper imaging. Nat Biotechnol 29:715–716. doi:10.1038/nbt.1941 pmid:21822247
    OpenUrlCrossRefPubMed
  43. ↵
    1. Lecoq J,
    2. Savall J,
    3. Vucinic D,
    4. Grewe BF,
    5. Kim H,
    6. Li JZ,
    7. Kitch LJ,
    8. Schnitzer MJ
    (2014) Visualizing mammalian brain area interactions by dual-axis two-photon calcium imaging. Nat Neurosci 17:1825–1829. doi:10.1038/nn.3867 pmid:25402858
    OpenUrlCrossRefPubMed
  44. ↵
    1. Lin MZ,
    2. Schnitzer MJ
    (2016) Genetically encoded indicators of neuronal activity. Nat Neurosci 19:1142–1153. doi:10.1038/nn.4359 pmid:27571193
    OpenUrlCrossRefPubMed
  45. ↵
    1. Liu R,
    2. Ball N,
    3. Brockill J,
    4. Kuan L,
    5. Millman D,
    6. White C,
    7. Leon A,
    8. Williams D,
    9. Nishiwaki S,
    10. de Vries S,
    11. Larkin J,
    12. Sullivan D,
    13. Slaughterbeck C,
    14. Farrell C,
    15. Saggau P
    (2018) Multi-plane imaging of neural activity from the mammalian brain using a fast-switching liquid crystal spatial light modulator. bioRxiv. Available at https://europepmc.org/abstract/PPR/PPR66128.
  46. ↵
    1. Lu R,
    2. Sun W,
    3. Liang Y,
    4. Kerlin A,
    5. Bierfeld J,
    6. Seelig JD,
    7. Wilson DE,
    8. Scholl B,
    9. Mohar B,
    10. Tanimoto M,
    11. Koyama M,
    12. Fitzpatrick D,
    13. Orger MB,
    14. Ji N
    (2017) Video-rate volumetric functional imaging of the brain at synaptic resolution. Nat Neurosci 20:620–628. doi:10.1038/nn.4516 pmid:28250408
    OpenUrlCrossRefPubMed
  47. ↵
    1. Madisen L,
    2. Garner AR,
    3. Shimaoka D,
    4. Chuong AS,
    5. Klapoetke NC,
    6. Li L,
    7. van der Bourg A,
    8. Niino Y,
    9. Egolf L,
    10. Monetti C,
    11. Gu H,
    12. Mills M,
    13. Cheng A,
    14. Tasic B,
    15. Nguyen TN,
    16. Sunkin SM,
    17. Benucci A,
    18. Nagy A,
    19. Miyawaki A,
    20. Helmchen F, et al
    . (2015) Transgenic mice for intersectional targeting of neural sensors and effectors with high specificity and performance. Neuron 85:942–958. doi:10.1016/j.neuron.2015.02.022 pmid:25741722
    OpenUrlCrossRefPubMed
  48. ↵
    1. Miller DR,
    2. Hassan AM,
    3. Jarrett JW,
    4. Medina FA,
    5. Perillo EP,
    6. Hagan K,
    7. Shams Kazmi SM,
    8. Clark TA,
    9. Sullender CT,
    10. Jones TA,
    11. Zemelman BV,
    12. Dunn AK
    (2017) In vivo multiphoton imaging of a diverse array of fluorophores to investigate deep neurovascular structure. Biomed Opt Express 8:3470–3481. doi:10.1364/BOE.8.003470 pmid:28717582
    OpenUrlCrossRefPubMed
  49. ↵
    1. Mukamel EA,
    2. Nimmerjahn A,
    3. Schnitzer MJ
    (2009) Automated analysis of cellular signals from large-scale calcium imaging data. Neuron 63:747–760. doi:10.1016/j.neuron.2009.08.009 pmid:19778505
    OpenUrlCrossRefPubMed
  50. ↵
    1. Nadella KM,
    2. Ros H,
    3. Baragli C,
    4. Griffiths VA,
    5. Konstantinou G,
    6. Koimtzis T,
    7. Evans GJ,
    8. Kirkby PA,
    9. Silver RA
    (2016) Random-access scanning microscopy for 3D imaging in awake behaving animals. Nat Methods 13:1001–1004. doi:10.1038/nmeth.4033 pmid:27749836
    OpenUrlCrossRefPubMed
  51. ↵
    1. Nimmerjahn A,
    2. Kirchhoff F,
    3. Helmchen F
    (2005) Resting microglial cells are highly dynamic surveillants of brain parenchyma in vivo. Science 308:1314–1318. doi:10.1126/science.1110647 pmid:15831717
    OpenUrlAbstract/FREE Full Text
  52. ↵
    1. Nimmerjahn A,
    2. Mukamel EA,
    3. Schnitzer MJ
    (2009) Motor behavior activates Bergmann glial networks. Neuron 62:400–412. doi:10.1016/j.neuron.2009.03.019 pmid:19447095
    OpenUrlCrossRefPubMed
  53. ↵
    1. Oh SW,
    2. Harris JA,
    3. Ng L,
    4. Winslow B,
    5. Cain N,
    6. Mihalas S,
    7. Wang Q,
    8. Lau C,
    9. Kuan L,
    10. Henry AM,
    11. Mortrud MT,
    12. Ouellette B,
    13. Nguyen TN,
    14. Sorensen SA,
    15. Slaughterbeck CR,
    16. Wakeman W,
    17. Li Y,
    18. Feng D,
    19. Ho A,
    20. Nicholas E, et al
    . (2014) A mesoscale connectome of the mouse brain. Nature 508:207–214. doi:10.1038/nature13186 pmid:24695228
    OpenUrlCrossRefPubMed
  54. ↵
    1. Oheim M,
    2. Beaurepaire E,
    3. Chaigneau E,
    4. Mertz J,
    5. Charpak S
    (2001) Two-photon microscopy in brain tissue: parameters influencing the imaging depth. J Neurosci Methods 111:29–37. doi:10.1016/S0165-0270(01)00438-1 pmid:11574117
    OpenUrlCrossRefPubMed
  55. ↵
    1. Ohki K,
    2. Chung S,
    3. Ch'ng YH,
    4. Kara P,
    5. Reid RC
    (2005) Functional imaging with cellular resolution reveals precise micro-architecture in visual cortex. Nature 433:597–603. doi:10.1038/nature03274 pmid:15660108
    OpenUrlCrossRefPubMed
  56. ↵
    1. Ouzounov DG,
    2. Wang T,
    3. Wang M,
    4. Feng DD,
    5. Horton NG,
    6. Cruz-Hernández JC,
    7. Cheng YT,
    8. Reimer J,
    9. Tolias AS,
    10. Nishimura N,
    11. Xu C
    (2017) In vivo three-photon imaging of activity of GCaMP6-labeled neurons deep in intact mouse brain. Nat Methods 14:388–390. doi:10.1038/nmeth.4183 pmid:28218900
    OpenUrlCrossRefPubMed
    1. Pachitariu M,
    2. Stringer C,
    3. Dipoppa M,
    4. Schröder S,
    5. Rossi LF,
    6. Dalgleish H,
    7. Carandini M,
    8. Harris KD
    (2017) Suite2p: beyond 10,000 neurons with standard two-photon microscopy. bioRxiv. Advance online publication. Retrieved July 20, 2017. doi: 10.1101/061507.
    OpenUrlAbstract/FREE Full Text
  57. ↵
    1. Piyawattanametha W,
    2. Barretto RP,
    3. Ko TH,
    4. Flusberg BA,
    5. Cocker ED,
    6. Ra H,
    7. Lee D,
    8. Solgaard O,
    9. Schnitzer MJ
    (2006) Fast-scanning two-photon fluorescence imaging based on a microelectromechanical systems two- dimensional scanning mirror. Opt Lett 31:2018–2020. doi:10.1364/OL.31.002018 pmid:16770418
    OpenUrlCrossRefPubMed
  58. ↵
    1. Pnevmatikakis EA,
    2. Soudry D,
    3. Gao Y,
    4. Machado TA,
    5. Merel J,
    6. Pfau D,
    7. Reardon T,
    8. Mu Y,
    9. Lacefield C,
    10. Yang W,
    11. Ahrens M,
    12. Bruno R,
    13. Jessell TM,
    14. Peterka DS,
    15. Yuste R,
    16. Paninski L
    (2016) Simultaneous denoising, deconvolution, and demixing of calcium imaging data. Neuron 89:285–299. doi:10.1016/j.neuron.2015.11.037 pmid:26774160
    OpenUrlCrossRefPubMed
  59. ↵
    1. Podgorski K,
    2. Ranganathan G
    (2016) Brain heating induced by near-infrared lasers during multiphoton microscopy. J Neurophysiol 116:1012–1023. doi:10.1152/jn.00275.2016 pmid:27281749
    OpenUrlCrossRefPubMed
  60. ↵
    1. Prevedel R,
    2. Verhoef AJ,
    3. Pernía-Andrade AJ,
    4. Weisenburger S,
    5. Huang BS,
    6. Nöbauer T,
    7. Fernández A,
    8. Delcour JE,
    9. Golshani P,
    10. Baltuska A,
    11. Vaziri A
    (2016) Fast volumetric calcium imaging across multiple cortical layers using sculpted light. Nat Methods 13:1021–1028. doi:10.1038/nmeth.4040 pmid:27798612
    OpenUrlCrossRefPubMed
  61. ↵
    1. Renshaw B,
    2. Forbes A,
    3. Morison BR
    (1940) Activity of isocortex and hippocampus: electrical studies with micro-electrodes. J Neurophysiol 3:74–105. doi:10.1152/jn.1940.3.1.74
    OpenUrlCrossRef
  62. ↵
    1. Rodríguez C,
    2. Liang Y,
    3. Lu R,
    4. Ji N
    (2018) Three-photon fluorescence microscopy with an axially elongated Bessel focus. Opt Lett 43:1914–1917. doi:10.1364/OL.43.001914 pmid:29652397
    OpenUrlCrossRefPubMed
  63. ↵
    1. Sofroniew NJ,
    2. Flickinger D,
    3. King J,
    4. Svoboda K
    (2016) A large field of view two-photon mesoscope with subcellular resolution for in vivo imaging. Elife 5:e14472. doi:10.7554/eLife.14472 pmid:27300105
    OpenUrlCrossRefPubMed
  64. ↵
    1. St Pierre F
    (2017) Two-photon imaging of neuronal voltage dynamics using genetically encoded voltage indicators. In: Optics in the Life Sciences Congress. Washington, DC: OSA.
  65. ↵
    1. Stirman JN,
    2. Smith IT,
    3. Kudenov MW,
    4. Smith SL
    (2016) Wide field-of-view, multi-region, two-photon imaging of neuronal activity in the mammalian brain. Nat Biotechnol 34:857–862. doi:10.1038/nbt.3594 pmid:27347754
    OpenUrlCrossRefPubMed
  66. ↵
    1. Stosiek C,
    2. Garaschuk O,
    3. Holthoff K,
    4. Konnerth A
    (2003) In vivo two-photon calcium imaging of neuronal networks. Proc Natl Acad Sci U S A 100:7319–7324. doi:10.1073/pnas.1232232100 pmid:12777621
    OpenUrlAbstract/FREE Full Text
  67. ↵
    1. Striker G,
    2. Subramaniam V,
    3. Seidel CA,
    4. Volkmer A
    (1999) Photochromicity and fluorescence lifetimes of green fluorescent protein. J Phys Chem B 103:8612–8617. doi:10.1021/jp991425e
    OpenUrlCrossRef
  68. ↵
    1. Svoboda K,
    2. Denk W,
    3. Kleinfeld D,
    4. Tank DW
    (1997) In vivo dendritic calcium dynamics in neocortical pyramidal neurons. Nature 385:161–165. doi:10.1038/385161a0 pmid:8990119
    OpenUrlCrossRefPubMed
  69. ↵
    1. Szalay G,
    2. Judák L,
    3. Katona G,
    4. Ócsai K,
    5. Juhász G,
    6. Veress M,
    7. Szadai Z,
    8. Fehér A,
    9. Tompa T,
    10. Chiovini B,
    11. Maák P,
    12. Rózsa B
    (2016) Fast 3D imaging of spine, dendritic, and neuronal assemblies in behaving animals. Neuron 92:723–738. doi:10.1016/j.neuron.2016.10.002 pmid:27773582
    OpenUrlCrossRefPubMed
  70. ↵
    1. Terada SI,
    2. Kobayashi K,
    3. Ohkura M,
    4. Nakai J,
    5. Matsuzaki M
    (2018) Super-wide-field two-photon imaging with a micro-optical device moving in post-objective space. Nat Commun 9:3550. doi:10.1038/s41467-018-06058-8 pmid:30177699
    OpenUrlCrossRefPubMed
  71. ↵
    1. Theer P,
    2. Denk W
    (2006) On the fundamental imaging-depth limit in two-photon microscopy. J Opt Soc Am A Opt Image Sci Vis 23:3139–3149. doi:10.1364/JOSAA.23.003139 pmid:17106469
    OpenUrlCrossRefPubMed
  72. ↵
    1. Trachtenberg JT,
    2. Chen BE,
    3. Knott GW,
    4. Feng G,
    5. Sanes JR,
    6. Welker E,
    7. Svoboda K
    (2002) Long-term in vivo imaging of experience-dependent synaptic plasticity in adult cortex. Nature 420:788–794. doi:10.1038/nature01273 pmid:12490942
    OpenUrlCrossRefPubMed
  73. ↵
    1. Tsai PS,
    2. Mateo C,
    3. Field JJ,
    4. Schaffer CB,
    5. Anderson ME,
    6. Kleinfeld D
    (2015) Ultra-large field-of-view two-photon microscopy. Opt Express 23:13833–13847. doi:10.1364/OE.23.013833 pmid:26072755
    OpenUrlCrossRefPubMed
  74. ↵
    1. Tsyboulski D,
    2. Orlova N,
    3. Saggau P
    (2017) Amplitude modulation of femtosecond laser pulses in the megahertz range for frequency-multiplexed two-photon imaging. Opt Express 25:9435–9442. doi:10.1364/OE.25.009435 pmid:28437905
    OpenUrlCrossRefPubMed
  75. ↵
    1. Tsyboulski D,
    2. Orlova N,
    3. Griffin F,
    4. Seid S,
    5. Lecoq J,
    6. Saggau P
    (2018) Remote focusing system for simultaneous dual-plane mesoscopic multiphoton imaging. bioRxiv. Advance online publication. Retrieved December 20, 2018. doi: 10.1101/503052.
    OpenUrlAbstract/FREE Full Text
  76. ↵
    1. Voigt FF,
    2. Chen JL,
    3. Krueppel R,
    4. Helmchen F
    (2015) A modular two-photon microscope for simultaneous imaging of distant cortical areas in vivo. In: Multiphoton microscopy in the biomedical sciences, Vol XV. SPIE BiOS, San Francisco, California.
  77. ↵
    1. Weisenburger S,
    2. Tejera F,
    3. Demas J,
    4. Chen B,
    5. Manley J,
    6. Sparks FT,
    7. Martínez Traub F,
    8. Daigle T,
    9. Zeng H,
    10. Losonczy A,
    11. Vaziri A
    (2019) Volumetric Ca2+ imaging in the mouse brain using hybrid multiplexed sculpted light microscopy. Cell 177:1050–1066.e14. doi:10.1016/j.cell.2019.03.011 pmid:30982596
    OpenUrlCrossRefPubMed
  78. ↵
    1. Wekselblatt JB,
    2. Flister ED,
    3. Piscopo DM,
    4. Niell CM
    (2016) Large-scale imaging of cortical dynamics during sensory perception and behavior. J Neurophysiol 115:2852–2866. doi:10.1152/jn.01056.2015 pmid:26912600
    OpenUrlCrossRefPubMed
  79. ↵
    1. Wilt BA,
    2. Fitzgerald JE,
    3. Schnitzer MJ
    (2013) Photon shot noise limits on optical detection of neuronal spikes and estimation of spike timing. Biophys J 104:51–62. doi:10.1016/j.bpj.2012.07.058 pmid:23332058
    OpenUrlCrossRefPubMed
  80. ↵
    1. Yang W,
    2. Miller JE,
    3. Carrillo-Reid L,
    4. Pnevmatikakis E,
    5. Paninski L,
    6. Yuste R,
    7. Peterka DS
    (2016) Simultaneous multi-plane imaging of neural circuits. Neuron 89:269–284. doi:10.1016/j.neuron.2015.12.012 pmid:26774159
    OpenUrlCrossRefPubMed
  81. ↵
    1. Yildirim M,
    2. Sugihara H,
    3. So PT,
    4. Sur M
    (2019) Functional imaging of visual cortical layers and subplate in awake mice with optimized three-photon microscopy. Nat Commun 10:177. doi:10.1038/s41467-018-08179-6 pmid:30635577
    OpenUrlCrossRefPubMed
  82. ↵
    1. Zhang S,
    2. Xu M,
    3. Kamigaki T,
    4. Hoang Do JP,
    5. Chang WC,
    6. Jenvay S,
    7. Miyamichi K,
    8. Luo L,
    9. Dan Y
    (2014) Selective attention. long-range and local circuits for top-down modulation of visual cortex processing. Science 345:660–665. doi:10.1126/science.1254126 pmid:25104383
    OpenUrlAbstract/FREE Full Text
  83. ↵
    1. Zong W,
    2. Wu R,
    3. Li M,
    4. Hu Y,
    5. Li Y,
    6. Li J,
    7. Rong H,
    8. Wu H,
    9. Xu Y,
    10. Lu Y,
    11. Jia H,
    12. Fan M,
    13. Zhou Z,
    14. Zhang Y,
    15. Wang A,
    16. Chen L,
    17. Cheng H
    (2017) Fast high-resolution miniature two-photon microscopy for brain imaging in freely behaving mice. Nat Methods 14:713–719. doi:10.1038/nmeth.4305 pmid:28553965
    OpenUrlCrossRefPubMed
Back to top

In this issue

The Journal of Neuroscience: 39 (46)
Journal of Neuroscience
Vol. 39, Issue 46
13 Nov 2019
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
  • Advertising (PDF)
  • Ed Board (PDF)
Email

Thank you for sharing this Journal of Neuroscience article.

NOTE: We request your email address only to inform the recipient that it was you who recommended this article, and that it is not junk mail. We do not retain these email addresses.

Enter multiple addresses on separate lines or separate them with commas.
Wide. Fast. Deep: Recent Advances in Multiphoton Microscopy of In Vivo Neuronal Activity
(Your Name) has forwarded a page to you from Journal of Neuroscience
(Your Name) thought you would be interested in this article in Journal of Neuroscience.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Print
View Full Page PDF
Citation Tools
Wide. Fast. Deep: Recent Advances in Multiphoton Microscopy of In Vivo Neuronal Activity
Jérôme Lecoq, Natalia Orlova, Benjamin F. Grewe
Journal of Neuroscience 13 November 2019, 39 (46) 9042-9052; DOI: 10.1523/JNEUROSCI.1527-18.2019

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Respond to this article
Request Permissions
Share
Wide. Fast. Deep: Recent Advances in Multiphoton Microscopy of In Vivo Neuronal Activity
Jérôme Lecoq, Natalia Orlova, Benjamin F. Grewe
Journal of Neuroscience 13 November 2019, 39 (46) 9042-9052; DOI: 10.1523/JNEUROSCI.1527-18.2019
Reddit logo Twitter logo Facebook logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Jump to section

  • Article
    • Abstract
    • Introduction
    • Discussion
    • Footnotes
    • References
  • Figures & Data
  • Info & Metrics
  • eLetters
  • PDF

Keywords

  • imaging
  • in vivo
  • multiphoton
  • two-photon

Responses to this article

Respond to this article

Jump to comment:

No eLetters have been published for this article.

Related Articles

Cited By...

More in this TOC Section

  • The Importance of Accounting for Movement When Relating Neuronal Activity to Sensory and Cognitive Processes
  • Characterizing Cortex-Wide Dynamics with Wide-Field Calcium Imaging
  • The Property-Based Practical Applications and Solutions of Genetically Encoded Acetylcholine and Monoamine Sensors
Show more TechSights
  • Home
  • Alerts
  • Visit Society for Neuroscience on Facebook
  • Follow Society for Neuroscience on Twitter
  • Follow Society for Neuroscience on LinkedIn
  • Visit Society for Neuroscience on Youtube
  • Follow our RSS feeds

Content

  • Early Release
  • Current Issue
  • Issue Archive
  • Collections

Information

  • For Authors
  • For Advertisers
  • For the Media
  • For Subscribers

About

  • About the Journal
  • Editorial Board
  • Privacy Policy
  • Contact
(JNeurosci logo)
(SfN logo)

Copyright © 2023 by the Society for Neuroscience.
JNeurosci Online ISSN: 1529-2401

The ideas and opinions expressed in JNeurosci do not necessarily reflect those of SfN or the JNeurosci Editorial Board. Publication of an advertisement or other product mention in JNeurosci should not be construed as an endorsement of the manufacturer’s claims. SfN does not assume any responsibility for any injury and/or damage to persons or property arising from or related to any use of any material contained in JNeurosci.