Image-based method for retrospective correction of physiological motion effects in fMRI: RETROICOR

Magn Reson Med. 2000 Jul;44(1):162-7. doi: 10.1002/1522-2594(200007)44:1<162::aid-mrm23>3.0.co;2-e.

Abstract

Respiration effects and cardiac pulsatility can induce signal modulations in functional MR image time series that increase noise and degrade the statistical significance of activation signals. A simple image-based correction method is described that does not have the limitations of k-space methods that preclude high spatial frequency correction. Low-order Fourier series are fit to the image data based on time of each image acquisition relative to the phase of the cardiac and respiratory cycles, monitored using a photoplethysmograph and pneumatic belt, respectively. The RETROICOR method is demonstrated using resting-state experiments on three subjects and compared with the k-space method. The method is found to perform well for both respiration- and cardiac-induced noise without imposing spatial filtering on the correction. Magn Reson Med 44:162-167, 2000.