Comparative test-retest reliability of metabolite values assessed with magnetic resonance spectroscopy of the brain. The LCModel versus the manufacturer software

Neurol Res. 2009 Jun;31(5):472-7. doi: 10.1179/174313209X395481. Epub 2009 Feb 12.

Abstract

Objective: Reproducibility is an essential strength of any diagnostic technique for cross-sectional and longitudinal works. To determine in vivo short-term comparatively, the test-retest reliability of magnetic resonance spectroscopy (MRS) of the brain was compared using the manufacturer's software package and the widely used linear combination of model (LCModel) technique.

Methods: Single-voxel H-MRS was performed in a series of patients with different pathologies on a 1.5 T clinical scanner. Four areas of the brain were explored with the point resolved spectroscopy technique acquisition mode; the echo time was 35 milliseconds and the repetition time was 2000 milliseconds. We enrolled 15 patients for every area, and the intra-individual variations of metabolites were studied in two consecutive scans without removing the patient from the scanner. Curve fitting and analysis of metabolites were made with the software of GE and the LCModel. Spectra non-fulfilling the minimum criteria of quality in relation to linewidths and signal/noise ratio were rejected.

Results: The intraclass correlation coefficients for the N-acetylaspartate/creatine (NAA/Cr) ratios were 0.93, 0.89, 0.9 and 0.8 for the posterior cingulate gyrus, occipital, prefrontal and temporal regions, respectively, with the GE software. For the LCModel, the coefficients were 0.9, 0.89, 0.87 and 0.84, respectively. For the absolute value of NAA, the GE software was also slightly more reproducible than LCModel. However, for the choline/Cr and myo-inositol/Cr ratios, the LCModel was more reliable than the GE software. The variability we have seen hovers around the percentages observed in previous reports (around 10% for the NAA/Cr ratios).

Conclusion: We did not find that the LCModel software is superior to the software of the manufacturer. Reproducibility of metabolite values relies more on the observance of the quality parameters than on the software used.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Amino Acids / metabolism
  • Brain / metabolism
  • Brain Diseases / diagnosis
  • Humans
  • Magnetic Resonance Spectroscopy / instrumentation*
  • Middle Aged
  • Organic Chemicals / metabolism
  • Reproducibility of Results
  • Software*

Substances

  • Amino Acids
  • Organic Chemicals